The US Food and Drug Administration has listed a Class I recall covering certain versions of three Abbott glucose-monitoring phone apps: FreeStyle Libre 3, Libre by Abbott and Lingo. The FDA posted the classification on 25 November 2026. Abbott began notifying customers in October. If you use one of these apps in the US, check for the latest version of the exact app you use.
This is a software correction. It does not tell people to remove their sensors simply because they use these apps. It is separate from the earlier FDA recall of certain FreeStyle Libre 3 and Libre 3 Plus sensors, which involved inaccurate low readings from affected physical sensors.
Which apps should I check?
Use the name on your app, rather than the name on your sensor box. The FDA records for this event list three app products. FreeStyle Libre 3 and Libre by Abbott are covered by both the security and signal-loss corrections. The Lingo app is covered by the security correction, but Abbott says the signal-loss problem described here does not affect Lingo.
Abbott’s US customer notice identifies FreeStyle Libre 3 version 3.6.10 and Libre by Abbott version 1.4.2 as releases that address both issues. The FDA Lingo record directs users to update and names version 1.12.3. These are the versions specified in the notices, not a reason to downgrade a newer installation. Get the latest version available from your phone’s official app store.
These are US instructions. The app names and available fixes can differ by country. The Canadian app notice is separate and should not be used to decide which US app version to install.
Why does the update matter?
For some FreeStyle Libre 3 and Libre by Abbott users, the signal could drop and the sensor might not reconnect to the phone app. When a phone app is not receiving readings, people who rely on it may miss a high or low glucose alert. The update also addresses a security weakness. A Class I designation reflects the seriousness of the potential risk; it does not mean that everyone with an affected version will experience a problem.
What can I do now?
- Open the App Store or Google Play on the phone you use with your sensor.
- Find your current app by its exact name and install the latest update offered. If your app already updated automatically, check its version rather than assuming it is current.
- Follow Abbott’s instructions to check the installed version. If the app still cannot connect, contact Abbott through the US customer notice for help with your particular app and phone.
Do not assume that no reading or alarm means your glucose is in range. If you rely on glucose readings for treatment decisions and the app is not providing them, follow your agreed backup monitoring plan. If a sensor reading does not match how you feel, follow your device instructions on confirming it with a blood glucose meter. The recall notice is about updating software, not changing insulin doses.
For background on how readings and alerts fit into everyday monitoring, see our FreeStyle Libre app guide.