Tandem Mobi Insulin Pump Recall: Check Your Serial Number

The US Tandem Mobi recall has different advice depending on your pump serial number. See which pumps need replacement and what the FDA says to do.

The US Food and Drug Administration has posted a recall notice for Tandem Mobi insulin pumps with Interoperable Technology. Some units need replacement. Other users of the named pump should follow updated instructions. The deciding detail is the pump’s serial number, not the phone operating system or simply the name on the pump.

The FDA listed the notice on 25 November 2026 and describes this as its most serious type of recall. Tandem sent its customer letters on 23 September. This is a separate safety action from the Tandem Mobi Android app recall.

Which pumps need to be replaced?

Open the FDA recall notice and follow its link to the current list of units marked for replacement. Compare the exact serial number on your pump or in your device information. Do not assume that a different serial number is affected just because it is close to one on the list.

If your number is listed: the notice says to request a return and replacement straight away through the link in Tandem’s customer letter or by calling US Tandem Customer Technical Support at 877-801-6901. Check the official instructions for using the pump while arranging replacement, and make sure you have the backup plan agreed with your diabetes team.

If your number is not listed: the FDA notice says you can continue to use your Tandem Mobi while following the updated instructions. Repeated or unusually frequent alarms that still interrupt delivery after you have followed the device guide are a reason to contact Tandem support.

Why is the pump sounding a blockage alarm?

An occlusion alarm means the pump has detected what it interprets as a blockage. Tandem reported that some Mobi pumps can signal a blockage even when there is none. The pump suspends insulin delivery when this alarm occurs. Delivery stays suspended until the user acknowledges the alarm and resumes therapy.

You cannot tell from the sound alone whether there is a real blockage. The FDA notice directs users to respond to each alarm and follow the troubleshooting instructions in their Tandem User Guide. This includes checking the cartridge, tubing and infusion site, correcting any problem found, and following the guide for resuming delivery. If you continue using the pump, the notice also recommends enabling the app’s push notifications so alerts can reach your phone.

Repeated interruptions can lead to high glucose and, in severe cases, the need for hospital care. The FDA notice says Tandem had reported 1,248 serious injuries associated with the issue as of 28 October 2026, with no deaths reported by that date. That is a dated report count, not an estimate of how often the problem happens to an individual user.

What should I do if delivery keeps stopping?

Keep the pump’s user instructions and your agreed backup plan to hand. The FDA notice recommends regular glucose checks for users awaiting replacement and preparing a backup method of insulin delivery as directed by your healthcare provider. If alarms keep interrupting treatment or you cannot restore delivery, contact Tandem and your diabetes team promptly. Follow your clinical team’s instructions for glucose or ketone checks and when to seek urgent help. Do not improvise insulin doses or stop insulin while waiting for a replacement.

This is a US notice. Users elsewhere should check the instructions for their country with Tandem or their local device supplier. Questions about the serial number, replacement or the alarm belong with Tandem; questions about backup insulin and high glucose belong with your diabetes care team.